(1 thru 7 from /local/data/gcn3a/array_cal/coarse_grid/)
(8 thru 10 from /local/data/gcn3c/array_cal/coarse_grid/)
example command:
batgse2dpi cg17_x_ba133_030304_19.list cg17_x_ba133_030304_19.dpi_window histmode="window" windows="/home/lhea/derek/windows/ba_133.window"
example command:
bathotpix cg17_x_ba133_030304_19.dpi_window cg17_x_ba133_030304_19.mask2_window chatter=2
example command:
batfftimage cg17_x_ba133_030304_19.dpi_window cg17_x_ba133_030304_19.img_window attitude=NONE detmask=cg17_x_ba133_030304_19.mask2_window bat_z=305.370972
(bat_z comes from Jay Cummings' analysis)
example command:
batcelldetect cg17_x_ba133_030304_19.img_window cg17_x_ba133_030304_19.src_window 6.0
example command:
batclean cg17_x_ba133_030304_19.dpi_window cg17_x_ba133_030304_19.dpi_window_clean cg17_x_ba133_030304_19.src detmask=cg17_x_ba133_030304_19.mask2_window srcclean=YES outversion="bkgcleaned"
By default, counts in the dpi are dead-time corrected (that is, they are multiplied by exposure/live-time).

C: dead-time corrected counts from windowed dpi
(counts*exposure/live_time)
t: exposure time for the data set
(See the IDL routine used to generate the measured count rate for each detector) I examined the measured count rates in two ways:
S: The number of photons/s emitted by the source into 4π
Source used: Ba-133-037
According to Nadine’s calibration report (calibhigh.xls):
S31 = 2.96 x 107 photons/s (on 6/27/03)
on 3/4/03, this rate would have been higher by a factor of (1/2)^(-115/3839)
(=1.02), making it 3.02 x 107 photons/s
on 7/1/03, this rate would have been lower by a factor of (1/2)^(4/3839)
(=0.999), making it 2.96 x 107 photons/s
S35 = 6.53 x 106 photons/s (on 6/27/03)
on 3/4/03, this rate would have been higher by a factor of (1/2)^(-115/3839)
(=1.02), making it 6.66 x 106 photons/s
on 7/1/03, this rate would have been lower by a factor of (1/2)^(4/3839)
(=0.999), making it 6.52 x 106 photons/s
S53 = 5.53 x 105 photons/s (on 6/27/03)
on 3/4/03, this rate would have been higher by a factor of (1/2)^(-115/3839)
(=1.02), making it 5.65 x 105 photons/s
on 7/1/03, this rate would have been lower by a factor of (1/2)^(4/3839)
(=0.999), making it 5.53 x 105 photons/s
S80 = 9.85 x 106 photons/s (on 6/27/03)
on 3/4/03, this rate would have been higher by a factor of (1/2)^(-115/3839)
(=1.02), making it 1.01 x 107 photons/s
on 7/1/03, this rate would have been lower by a factor of (1/2)^(4/3839)
(=0.999), making it 9.84 x 106 photons/s
r: distance from the source to the detector
Photons/s/cm2 incident on a fully-illuminated detector at a
distance r from the source.
fatten: fraction of photons that are transmitted through the passive materials between the source and the detectors (not including the lead tiles)
fillum:
fraction of photons that are transmitted through the mask
Aeff: Effective area of the detector
The Cosine Correction Factor is:
QE = the quantum efficiency of a detector at a particular photon energy and at a particular angle
I then compared the results of the measured rates with the predicted rates.
Plotted here are histograms of the measured count rates and predicted count rates from all of the "good" detectors in each run. Only center detectors are included in the histograms.
black: histogram of measured rates
red: histogram of predicted rates
The predicted rates are all systematically higher than the measured rates. After the plots, there is a table that compares these peak rates along with the mask-weighted summed rates.
back-of-envelope calculation:
The source is at x=-121.18, y=3.84, z=304.8 (all in cm). A fully-illuminated detector as close as possible to the source would be at x=-59.85, y=3.84, z=304.8, making r=310.9 cm. The fraction of photons transmitted through all passive materials (when the source is on axis), including air is 0.813 (31 keV photons), 0.836 (35 keV photons), 0.877 (53 keV photons), and 0.893 (80 keV photons). Also, for a detector at that position, QE31 = QE35 = QE53 = 1, and QE80 = 0.98, and the cosine correction factor is approximately 1. fillum for 31, 35, and 53 keV photons is 1. For 80 keV photons, it's 0.948. For 31 keV photons, 86.5% of counts are above the 17 keV threshold. For 35 keV photons, 79.1% of counts are above the 17 keV threshold. Overall, this would give a count rate of:










The "Max Measured Rate" and the "Max Predicted Rate" columns in this table show the count rates that correspond to the eyeball-middle of the high-rate peak in the histograms.
|
Run ID |
tan(θ) |
Summed Rate |
Summed Rate |
Ratio |
Max Measured |
Max Predicted |
Ratio |
|---|---|---|---|---|---|---|---|
|
3 |
0.142 |
4.22 |
4.37 |
0.97 |
4.4 |
4.6 |
0.96 |
|
4 |
0.179 |
4.15 |
4.33 |
0.96 |
4.3 |
4.6 |
0.93 |
|
33 |
0.252 |
4.31 |
4.10 |
1.05 |
4.6 |
4.3 |
1.07 |
|
34 |
0.305 |
3.66 |
3.79 |
0.96 |
4.1 |
4.5 |
0.91 |
|
17 |
0.370 |
3.31 |
3.44 |
0.96 |
3.4 |
4.4 |
0.77 |
|
18 |
0.444 |
2.90 |
3.00 |
0.96 |
3.1 |
3.0 |
1.03 |
|
25 |
0.555 |
1.83 |
2.02 |
0.91 |
2.5 |
2.7 |
0.93 |
|
88 |
0.559 |
1.21 |
1.83 |
0.66 |
2.2 |
2.4 |
0.92 |
|
87 |
0.595 |
1.48 |
1.71 |
0.87 |
2.1 |
2.2 |
0.95 |
|
82 |
0.727 |
0.72 |
0.79 |
0.91 |
1.1 |
1.2 |
0.92 |
There is a lot of scatter in these measurements but,
The maximum measured rates are (on average) 6.1% lower than the maximum
predicted rates.
The summed measured rates are (on average) 7.9% lower than the summed
predicted rates.
The 133Ba runs produce slightly lower rates than expected.
It may be that some of the events from the dominant 31 keV line are partially cut off by the electronics threshold. An analysis of just the 80 keV and 53 keV lines, or (better yet) a fit of the 80 and 53 keV peaks to the response matrix would tell whether this was the case.
Changes from 11 Jun 2004:
Changes from 26 Feb 2004:
Changes from 5 Feb 2004:
Changes from 15 Dec 2003:
Return to main BAT response page
Return to Home Page of Derek Hullinger