Method of Finding μτ Products for Electrons and Holes

Derek Hullinger
8 March 2002

Rev. C (18 April 2002):

Properly referenced Goro Sato's work
Added a note about validity of eq. 2 (Rev. B)

1. Introduction

The purpose of this experiment was to test a method for finding the μτ product for electrons and for holes using two spectra taken at different bias voltages. The use of two bias voltages for finding the μτ product for electrons was proposed to me by Goro Sato of Tokyo University, who used a similar, though more sophisticated, technique to determine μτ products for CZT detectors.1 The analysis requires:

  1. A pulser sweep
  2. A low bias voltage spectrum
  3. A high bias voltage spectrum

When I first began, all three of these data sets weren't yet available on any test sandwiches, so I took measurements on a single-detector setup (similar to the setup used by Goro Sato) using the Canberra system in Bert's lab.

At first, I made measurements using a simple setup involving only a source, a detector, and the eV Products brass detector holder. I found that there were scattering features in the spectrum that weren't consistent with a simple model of just the holder and source. In order to control these features so that I could properly model them with grmcflight, I determined to shield the entire setup.

2. Experimental Procedures

I placed a Cobalt-57 source approximately 13 cm from the detector holder. The entire setup was enclosed inside 2 mm thick lead shielding with a small circular opening for the end of the brass detector holder and a small square hole behind the source (see figure 1). The purpose of the small square hole was to reduce backscattering--photons that would have backscattered (if the hole had not been present) would instead exit the setup, and the probability that they would scatter off of outside materials and re-emerge through the hole would be small.

Figure 1: Test Setup

Drawing of Test Setup

The gain was adjusted so that a 150 keV range covered approximately 4000 channels. The LLD was adjusted to just eliminate the noise spike at the low end of the spectrum. Two spectra were taken, each for 1 hour (live time): one at 300 V bias voltage and one at 100 V bias voltage. Then a pulser sweep was recorded at 7 positions spanning the 4000 channel range.

Figure 2: Spectra Taken with 100 V and 300 V Bias Voltages

Plots of two spectra, which resemble each other except that the
peaks of one are shifted to the left of those of the other

When the bias voltage was set to 100 V, the spectrum shifted to lower channels.

3. Background

The Hecht Relation describes the charge trapping effects of a detector. It gives the charge collection efficiency as a function of depth of interaction:

Equation which gives the charge collection efficiency(1)
where:
Q is the charge collected at the electrodes of the detector,
Q0 is the total charge generated at the interaction site,
μ is the mobility of a charge carrier in the detector,
τ is the average lifetime of a charge carrier in the detector,
(μτ)e is the μτ product for electrons,
(μτ)h is the μτ product for holes,
D is the thickness of the detector (2 mm),
Vb is the bias voltage,
z is the depth of interaction

Figure 3 shows charge collection efficiency vs. depth of interaction for some typical values of Vb, (μτ)e, and (μτ)h:

Figure 3: The Hecht Relation

Hecht Relation plotted against depth

Because of this effect, photons that interact deeper within the detector are "measured" as having deposited less energy. The overall effect is that photons of a given energy produce a peak having a tail that extends to lower energies.

4. Finding μτ for Electrons

Charge collected is related to channel by:

charge collection efficiency as a function of channel(2)

where:
Vinduced is the height of the voltage pulse induced by the charge resulting from photons interacting at a particular depth,
VMax is the height of the voltage pulse that would have been induced if all the charge had been collected,
Ch is the channel that counts are recorded in when Vinduced is measured,
ChMax is the channel that all counts would have been recorded in if all the charge had been collected,
A is the electronics gain,
O is the electronics offset.

Note that this equation only holds if the induced charge Q is proportional to the induced voltage pulse height Vinduced, and only if voltage pulse height V is linear with channel Ch.

4.1 Finding the Gain and Offset

For each pulser peak, I recorded the dial setting and measured the voltage height on an oscilloscope. Next I found the the channel positions of each pulser peak by fitting it to a gaussian. I fit a line to the voltage heights vs. the channel positions and thus obtained the electronics gain and offset.

4.2 Method

Photons interacting near the front surface of the detector (z=0) are the most numerous and result in the highest efficiencies, so they produce counts in the main, high-channel part of the peak. Setting z=0 in equation 1 produces:

Equation for the Peak Charge Collection Efficiency.(3)

Note that this equation is independent of (μτ)h.

If the position of a peak at one bias voltage and the position of the same peak at another bias voltage are determined, these values can be substituted into equation 3, along with the known values of A, O, and D, producing two equations with two unknowns: (μτ)e and ChMax.

4.3 Results

The positions of the 122 keV peak for the two spectra were:

Bias Voltage Peak Position
100 V 2602 ± 3
300 V 2730 ± 3

Solving equations 3 numerically for (μτ)e gives:

(μτ)e = (2.72 ± 0.13) x 10-3 cm2/V

5. Finding μτ for holes

5.1 Method

I wrote a routine, in IDL, which takes 6 parameters:

1) gain (in energy units),
2) offset (in energy units),
3) normalization,
4) (μτ)e,
5) (μτ)h, and
6) FWHM line resolution (in channel units),

and returns the spectrum that would result from a stream of incoming photons with a given energy and intensity, assuming uniform illumination and neglecting Compton scattering.

I have used this function in the past to fit spectra and try to obtain the best-fit (μτ)e and (μτ)h, but since (μτ)e, gain, and offset all affect the position of a peak, and gain and offset cannot be determined from a single spectrum, there were many possible pairs of (μτ)e and (μτ)h that, together with suitable values of the other 4 parameters, all fit a single peak equally well (without necessarily fitting the other lines in the spectrum well). Fitting an entire spectrum was problematic because the function does not include effects due to Compton scattering. Having an independent measure of (μτ)e helps.

Gain and offset had previously been determined in voltage units. To obtain them in energy units required a little algebra, involving ChMax found from the numerical solution of equation 3.

With (μτ)e known and fixed, I selected a 200 channel region surrounding the 122 keV peak of the 300 V spectrum (channels 2400 through 2800) and fit my IDL function to it. The fitting procedure I used did not converge to a local mininum very well (for some reason). Therefore, I used many different starting values for the parameters and recorded the set of best fit parameters and Χ2 for each fit.

5.2 Results

There is a cluster of values at (2.93 ± 0.13) x 10-5, but the values of (μτ)h with the minimum Χ2 are at (3.33± 0.17) x 10-5.

I Recognized that including more of the peak tail in the fit might lead to a more trustworthy value of (μτ)h, so I selected a 300 and 400 channel region and repeated the process. The results were quite different:

Minimum-Χ2 (μτ)h values

(10-5 cm2/V)

Channel Range (μτ)h
2600 - 2800 3.33±0.17
2500 - 2800 2.65±0.06
2400 - 2800 2.27±0.08

The plots of Χ2/n vs. (μτ)h for all 3 channel ranges are shown in Figure 4:

Figure 4: Χ2/n vs. (μτ)h

plot of reduced chi squared verses mu tau h

·: channels 2600 thru 2800
+: channels 2500 thru 2800
×: channels 2400 thru 2800

I plotted the real data verses the simulated spectra corresponding to best fit parameters at the center and extremes of each error bar. For example, for the channel range of 2600 to 2800, I used best fit parameters for which (μτ)h converged to 3.50, 3.67, and 3.84 x 10-5. These plots are shown in Figures 5 thru 7:

Figure 5: Data vs. Best Fit Plots

Channel Range 2600 thru 2800

plot of best fit curve for channels 2600 to 2800

Figure 6: Data vs. Best Fit Plots

Channel Range 2500 thru 2800

plot of best fit curve for channels 2500 to 2800

Figure 7: Data vs. Best Fit Plots

Channel Range 2400 thru 2800

plot of best fit curve for channels 2400 to 2800

Looking at the spectra, it is apparent that the parameters within each error bar produce spectra that are nearly identical. The plots of all 9 best fit spectra next to the real data show that the there are small differences between them (see Figures 8 and 9).

Figure 8: All 9 Best Fit Plots

plots of the 9 best fit curves verses the data

Figure 9: All 9 Best Fit Plots

(full channel range)

plots of the 9 best fit curves verses the data for the full channel
range

6. Conclusion

By taking a pulser sweep and spectra at 2 different bias voltages, I was able to obtain a value for (μτ)e that I feel confident with. But in trying to obtain a reliable value for (μτ)h, I found that the techniques I used led to some serious uncertainty.

7. (Near) Future Work

7.1 Fitting multiple peaks

It was suggested that the presence of the 136 keV line might be adding counts to the background of the 122 keV line, affecting the fitting, and that it would be wise to fit both the 122 keV and 136 keV lines. I have done this for channels 2400 thru 3100. I may not have time to put graphs together for them, but the there was a mean (μτ)h of 2.70 x 10-5 cm2/V with a standard deviation of 0.14 x 10-5 cm2/V.

An interesting note is that when both lines (122 keV and 136 keV) were used in the fit, the plot of Χ2/n vs. (μτ)h doesn't have a clear minimum trough as it did when only the 122 keV line was used in the fit.

7.2 Accuracy of (μτ)h values

Obviously, I need to be certain that the values I find are accurate. One good way to verify that I have found values that reproduce the spectra is to measure spectra and then model the setup of the measurement with grmcflight and see if the spectra can be reproduced. I have begun (but haven't finished) a model of the lead-shielded single-detector setup. I'm not sure how to model the test setup involving the sandwiches, but I understand that Ed and Ann are looking into how to modify the DAP "gd" (gosh-darn) component so it can be used for a sandwich or block.

7.3 Necessary Precision of (μτ)e and (μτ)h values

I don't have a full report ready yet, but I've looked into how precise the determination of (μτ)e and (μτ)h values need to be. I have looked into how changing (μτ)e and (μτ)h affect a line at 122 keV.

In the case of (μτ)e, a deviation of 0.1 x 10-3 cm2/V changes the the peak position by 0.2% or less (depending on the initial values of (μτ)e and (μτ)h), the peak height by 0.9% or less, and the tail height by 0.9% or less (I somewhat arbitrarily defined the tail height to be the height of the tail at a point 250 channels (10.75 keV) below the peak).

In the case of (μτ)h, a deviation of 0.1 x 10-5 cm2/V changes the the peak position by 0.03% or less, the peak height by 2.0% or less, and the tail height by 0.02% or less. (0.03% was about the size of my error due to the channel spacing.)

7.4 Simultaneous Fit of both 100 V and 300 V measurements

I believe I've figured out how to go about this. This is the technique used by the ISAS group. I also plan, to check the accuracy, to use more than one pair of voltages.

1See Goro Sato, Tadayuki Takahashi, Masahiko Sugiho, Manabu Kouda, Shin Watanabe, Yuu Okada, Takefumi Mitani, and Kazuhiro Nakazawa, Characterization of CdTe/CdZnTe detectors, submitted to IEEE Transactions on Nuclear Science 2001

Return to Home Page