Comparison of Simulation Data
with Calibration Data

Derek Hullinger
7 Jan 2004

Bottom Line:

The number of counts in the calibration data is about 24% higher than the number of counts in the simulated data.

However, the predicted number of counts (from calculation) is only about 8% higher than the number of counts in the simulated data.

(The values and histogram plots given below were generated with this IDL routine.)

Comparison of Spectra:

Results:



Calibration Data
Calibration Data (Scaled to 122 keV peak of Sim Data)
Simulation Data

At the 122 keV peak, the simulation data is about 30% lower than the calibration data.

Comparison of Total Counts (above 20 keV)

Results:

Total Simulation Counts (above 20 keV):
14.2 counts
Total Calibration Counts (above 20 keV):
18.7 counts

Total Predicted Counts (above 20 keV):
15.5 counts

The total counts in the simulation is 24% lower than the total counts in the calibration data. This is consistent with what is seen in the plot.

The total predicted counts falls in between the simulation result and the calibration result, being 18% lower than the total calibration counts and is 8% higher than the total simulation counts.

Back-of-Envelope Calculation:

The source is located at x=-7.769216, y=4.365952, and z=294.4 (all in cm). A fully illuminated detector directly below the source would be at r=294.4 cm. The absorption efficiency in CZT of 122 keV photons is 0.65, and the absorption efficiency in CZT of 136 keV photons is 0.55. The cosine correction factor is 1. Using a source photon rate of 1.41 x 105 122 keV photons/s and 1.79 x 104 136 keV photons/s (where were the rates on 8/18/03), the predicted counts is: [(1.41 x 105)*/(4*π*294.4*294.4)*(0.16)*(0.65)+
(1.79 x 104)*/(4*π*294.4*294.4)*(0.16)*(0.55)]
*1073
=16.0 counts.

This agrees very well (within 3%) with the predicted total counts above.

Comparison of Histograms:


Simulation Data
Calibration Data
Predicted Data

The peak positions in these background-subtracted histograms match the totals above quite nicely.

The peak of the simulation data histogram is not as well-defined, due to the low statistics. The total counts found by sum_spectra is a little higher than the position of the peak, but this might be simply because of the low statistics.

Conclusions:

The difference between the simulation and the calibration data is large, but it is interesting that the difference between simulation and prediction is not so large--only 8%. Any materials included in the simulation (other than the lead tiles) that lie between the source and the detectors would attenuate the beam and decrease the total number of counts. Could 8% of the photons be removed by scattering? If so, that could explain the difference. There's still the concern over why the calibration data gives a much larger value, though. That, I don't have an explanation for yet.

Return to main BAT response page