Comparison of Simulation Data
with Calibration Data

Derek Hullinger
6 Feb 2004

I. Bottom Line:

The number of counts in the calibration data is about 42% higher than the number of counts in the simulated data.

However, the predicted number of counts (from calculation) agrees with the simulation to within 3%.

(The values and histogram plots given below were generated with this IDL routine.)

II. Comparison of Spectra:

Results:



Calibration Data
Calibration Data (Scaled to 122 keV peak of Sim Data)
Simulation Data

At the 122 keV peak, the calibration data is about 50% higher than the simulation data.

III. Comparison of Total Counts (above 20 keV)

Results:

Total Counts from Simulation (above 20 keV):
77.2 counts
Total Counts from Calibration Data (above 20 keV):
109.4 counts

Total Counts from Calculation (above 20 keV):
79.4 counts

The total counts from the calibration data is 42% higher than the total counts from the simulation. This is pretty consistent with what is seen in the plot.

The total calculated counts is within 3% of the total simulation counts.

IV. Back-of-Envelope Calculation:

The source is located at x=-7.778, y=4.129, and z=294.4 (all in cm). A fully illuminated detector directly below the source would be at r=294.4 cm. The absorption efficiency in CZT of 122 keV photons is 0.61, and the absorption efficiency in CZT of 136 keV photons is 0.49. The attenuation due to all of the other passive materials (except air) for 122 keV photons is 0.955, and for 136 keV photons is 0.956. Using a source photon rate of 8.25 x 105 122 keV photons/s and 1.045 x 105 136 keV photons/s, the predicted counts is: [(8.25 x 105)*/(4*π*294.4*294.4)*(0.955)*(0.16)*(0.61)+
(1.045 x 105)*/(4*π*294.4*294.4)*(0.956)*(0.16)*(0.49)]
*1073
=83.5 counts.

This agrees fairly very well with the predicted total counts above.

V. Comparison of Histograms:


Simulation Data
Calibration Data
Predicted Data

The peak positions in these background-subtracted histograms match the totals above very well.

VI. Conclusions:

The difference between the simulation and the calibration data is large, but the simulation matches the prediction very well—within 3%. Since all background (i.e., scattered) photons were removed by the background subtraction, the only reasons that the measurement would be higher than the simulation (that I have thought of so far) are:

Return to main BAT response page