Lower Threshold Study:
Variation between detector channels

Goro Sato
25 Oct 2004
Modifiled: 07 Dec 2004
Added a section "I-c)" to show the results for all the 16 blocks.

We investigated the lower threshold behavior. Note that the threshold can be changed through the input parameter of threshold in each different run at the level of sandwich. However, it is still worth investigating the overall behaviour using some specific runs. Here we examine the Co-57 block calibraion runs because there are high statistics in the Co-57 runs and the 14.4 keV line is just around lower threshold.


Data sets examined:

Co-57 block calibration runs:

Array NumberFile NameInput Threshold Value*
0blk_3_co57_200_20_0_021120_1.fits?
1blk_4_co57_200_20_0_021119_1.fits12 mV
2blk_8_co57_200_20_0_0_021209_1.fits12 mV
3blk_9_co57_200_20_0_0_021204_1.fits12 mV
4blk_13_co57_200_20_0_0_021218_1.fits12 mV
5blk_14_co57_200_20_0_0_021219_1.fits12 mV
6blk_11_co57_200_20_0_0_021227_1.fits12 mV
7blk_12_co57_200_20_0_0_021224_1.fits12 mV
8blk_a_co57_200_20_0_021111_1.fits?
9blk_17_co57_200_20_30_0_030205_3.fits12 mV
10blk_5_co57_200_20_0_0_021201_1.fits12 mV
11blk_6_co57_200_20_0_0_021126_3.fits12 mV
12blk_10_co57_200_20_0_0_021210_1.fits12 mV
13blk_7_co57_200_20_0_0_021211_1.fits12 mV
14blk_16_co57_200_0_0_021217_1.fits20 mV ?**
15blk_15_co57_200_20_0_0_021218_1.fits12 mV
*: The values are from the "block-level science calibration record sheets" (white/blue log binders stored in the BAT lab.)
**: It was noticed that the value was set to 20 mV after several runs.




I) Measured lower threshold


I-a) Method:

For each of the above data sets:

  1. corrected for the gain, offset and non-linearity for the electronics
  2. rebinned into 4096 ch for the range of 3000 mV in pulser voltage
  3. (The 11 testpulse peak and DNL run data were used to convert the raw ADU channel to pulser voltage unit.)

  4. found the first five continuous non-zero bins for each detector channel
    LD: the left edge of the five bins was assumed as the lower threshold



I-b) Results for Block 13:

Measured lower threshold (LD) in pulser voltage unit:

The LDs were extracted for each detector channel for all the 16 blocks.
Here we show the results for the 2048 detector channels in block 13.
The guidelines are drawn to separate the 16 sandwiches.

In this specific run, the "threshold" value in the operation panel was set to 12 mV.

This value (12 mV) is far from the actual LD. (Jay has already indicated this difference between the setup value and the pulser voltage.)

There is a variation in LD even within the same sandwiches.

Measured lower threshold (LD) in energy

Converted the pulser voltage into energy utilizing the 122 keV line position




I-c) Results for 16 Blocks:

Measured lower threshold (LD) in energy

II) Combined LD distribution for all the 16 blocks:

Although the data sets were acquired in separate operations, it is interesting to combine the LD distributions for the 16 blocks and to see how the LD distribuetes over the entire CZT array

LDs are ranged from less than 5 keV to over 15 keV.

Here, let us integrate the above histogram.

This "integrated" LD distribution enables us to know the number of detectors above threshold at each energy, or the number of "effective detectors" at each energy.
With 12 mV of the setup threshold over the 16 blocks, we can achieve ~97% of the detectors avobe the threshold at 15 keV.

III) Conclusions:

    1. I confirmed that there is a difference between the actual lower threshold in puser voltage and the setup value of the "threshold" in the operation panel.
    2. I found a large variation in lower threshold between detector channels more than 10 keV.
    3. However, if we can keep these threshold levels (12 mV), the nominal threshold of BAT instrument "15 keV" will be mostly achieved, which is good news!
    4. Below ~15 keV, the LD variation should be taken into account for the energy response (effective area).

Return to main BAT response page