MOXE Detector Efficiency

Soren asked whether we could get an idea of whether the efficiency of the detectors is a function of position from the current calibration data. I took the positioner test data, and compared the number of hits received as a function of angle with a simple model for what would be expected. The model assumes the same number of photons No was sent in during each run of the positioner, and that the angular dependence of the absorption is simply related to the thickness of the window as seen by the beam. This model does not take into account the variation of the shape of the PSF as a function of angle. For each image, I plot here the number of hits versus position, the model for each detector (the absorption coeffiecient and number of incident photons are calculated once for each detector by fitting to the data at the center and edges) and the percentage difference between the model and the data. Also shown is the histogram of the percentage difference.

  • Images and Histograms