![]() |
The spectra of most X-ray binaries may be described by a power law with an
exponential roll off above some cutoff energy Ec (e.g. White et al. (1983)).
The cutoff energy is
generally in the range 10-20 keV (Nagase et al. (1992)). The present data do not
require such a roll off. This indicates that Ec is greater than
the upper limit of BBXRT's energy range (12 keV), in
agreement with previous observations (Nagase et al. (1992)). The best fit
model is an absorbed power law continuum, typical of high mass X-ray
binaries, and emission at
6.67
0.07 keV (modeled using a
Gaussian). The
centroid of the emission line corresponds to transitions of
helium-like iron. The photon index is
1.10
0.02, the absorption
column
NH = (1.28
0.04) x 1022 cm-2, and the iron
emission line has equivalent width
198
45 keV and FWHM
740
210 keV. Fits to the A0 data are summarized in Table 1.
If the continuum is modeled by a simple absorbed power law there are
residuals of about 10% between 1 and 3 keV. They lie close in energy
to the silicon K and gold
MIV and MV edges, respectively. Similar residuals are present, but are less
significant, in the B0 data because of the poorer counting statistics.
It was found that these
could be removed from the A0 data by including two more components in the
model: Gaussian absorption
at
1.5
0.1 keV and gaussian emission at
2.24
0.04 keV. These have equivalent widths of 69
43 and
16
9 keV respectively. Taken together, these features are significant at the
84% confidence level. When they are included in the model the value
obtained for the column density, NH, is slightly lower, changing from
1.28 to
1.17 x 1022 cm-2. A similar correction to the
BBXRT response was
necessary for another bright source, Cyg X-2 (Smale et al. (1993)).
Fits to spectra of the Crab obtained with BBXRT have residuals
around 1.7 and 2.2 keV due to
uncertainties in modeling the Au and Si edges (Weaver et al. (1995)).
Owens et al. (1997) have shown that these features in the Crab spectrum
are due to
X-ray absorption fine structure and are a consequence of the high
spectral resolution of BBXRT combined with the brightness of the
source. These features in the A0 data were thus interpreted as instrumental
effects. It must be emphasized that these residuals only appear in data
from bright sources such as the Crab. In the majority of BBXRT
observations the counting rates were much lower and these residuals were
not detectable.
In their analysis of the spectrum of Cyg X-1, Marshall et al. (1993)
dealt with these instrumental features by ignoring a narrow band of
channels around 1.7 and 2.2 keV. It was decided that
the same approach would be appropriate for the Cen X-3 data. Ignoring these channels lowered
significantly and did not appreciably change the continuum parameters.
This enabled us to obtain acceptable fits to the A0 data.
Analysis of the B0 data yielded similar results. The iron K emission
line energy obtained from the B0 data is slightly lower than that
obtained with the A0 data as shown in Table 1. However, the
model which was fit to A0 gives an acceptable
when it is
fit to the
B0 data and only the overall normalization is allowed to vary. The
residuals around the silicon and gold edges are not as large as in the A0
data and it is not necessary to remove them to obtain a satisfactory fit.
A simultaneous fit to the A0 and B0 data yields an emission line whose
FWHM is
0.78
0.18 keV with a centroid energy of
6.63
0.06 keV and equivalent width
200
40 keV.
The presence of residuals around 6.4 keV led us to include a second
emission line in the model fit to the A0 data.
The best-fit centroid energy for this line is
6.40
0.03 keV.
This would correspond to
fluorescence of iron in low ionization stages. Its equivalent width is
42
18 keV. The best fit value for its physical width is zero with
a 90% confidence
upper limit for one interesting parameter of
120 keV FWHM which is less
than the FWHM energy resolution. Inclusion of this line in the fit
pushes the energy of the broad line to
6.77
0.05 keV and changes
its physical and equivalent widths to
420
180 keV FWHM and
128
37 keV respectively. The fit is shown in
Figure 2.2. A two line simultaneous
fit to the A0 and B0 data yields similar results as shown in
Table 1. Although the 6.4 keV line is certainly
physically plausible, it is only significant at the 97.2% or 2.2
level in
the A0 data. A second line is not required to obtain a satisfactory fit to
the B0 data. This is consistent with the 50% lower counting rate in the
B0 pixel as compared with the A0 pixel. When this model is fit to the B0 data
the two lines become degenerate.
![]() |
| Detector | A0 | B0 | A0 and B0 |
| Model 1: Absorbed power law | |||
| NH (1022 cm-2) |
1.28 |
1.26 |
1.27 |
|
1.10 |
1.06 |
1.09 |
|
|
|
1.326/393 | 1.151/384 | 1.241/778 |
| Model 2: Absorbed Power Law with one emission line | |||
| NH (1022 cm-2) |
1.13 |
1.29 |
1.30 |
|
1.14 |
1.10 |
1.13 |
|
| E (keV) |
6.67 |
6.5 |
6.63 |
|
|
0.32 |
0.4 |
0.34 |
| EW (keV) | 198 |
210 |
200 |
|
|
1.086/389 | 1.049/381 | 1.071/775 |
| Model 3: Absorbed power law with two emission lines 2 | |||
| NH (1022 cm-2) |
1.31 |
... |
1.30 |
|
1.13 |
... |
1.12 |
|
| E1 (keV) |
6.77 |
... |
6.73 |
|
|
0.18 |
... |
0.25 |
| EW1 (keV) | 128 |
... | 142 |
| E2 (keV) |
6.40 |
... |
6.40 |
|
|
< 0.052 | ... | < 0.055 |
| EW2 (keV) | 42 |
... | 36 |
|
|
1.069/386 | ... | 1.066/770 |
| Model 4: Absorbed power law with narrow lines fixed at | |||
| NH (1022 cm-2) |
1.31 |
1.28 |
1.30 |
|
1.13 |
1.08 |
1.12 |
|
| EW6.4 (keV) | 53 |
54 |
53 |
| EW6.7 (keV) | 60 |
41 |
54 |
| EW6.93 (keV) | 41 |
< 38 | 33 |
| EW7.53 (keV) | 26 |
< 22 | 17 |
|
|
1.063/388 | 1.076/380 | 1.074/774 |
There is also a feature at 7.5 keV which corresponds to K
emission from neutral nickel. It is doubtful that this is astrophysical
in origin because there are residuals around 7.5 keV in
both central pixels when the Crab spectrum is fit (Weaver et al. (1995)).
It is possible that this line is due to fluorescence in the nickel
mesh of the detector's blocking filter. When this feature is fit by a narrow
line in the A0
data it is significant at the 2
level. It is not required by the
B0 data and when it is included in a fit to B0 an equivalent width
consistent with zero results.
Fitting a blend of three narrow lines to the iron emission feature instead
of a broad and a narrow line yields a
which is not
significantly different. Such a fit to the A0 data is shown in
Figure 3.4. It includes another narrow line which is
significant at the 2
confidence level to fit the suspected nickel
feature. The energies of these lines were fixed at 6.4, 6.7, 6.93, and
7.53 keV, corresponding to emission by un-ionized iron, helium- and
hydrogen-like iron and un-ionized nickel. Thus the data are consistent with
either a broad line or a blend of narrow lines with physically plausible
energies. In an attempt to distinguish between the two line and four line
models, the energy range was restricted to 5.5-8 keV and the data were
fitted with the continuum parameters held fixed. The continuum parameters
were those obtained from fits to the 1-12 keV data. F-tests (Bevington (1969)) were still
unable to distinguish between the two models.
An attempt was made to improve the counting statistics and uncertainties
by relaxing the data quality filter. Fits were made to data in which the
only events rejected were those that had the VLE, pulse-pulse,
or pixel-pixel flags set. The resulting light curve was flat as the
telescope passed through the SAA, indicating that contaminating events
were being rejected without significant loss of good data. This is
because most of the background events from the SAA were at low energies
and had already been rejected by restricting the analysis to energies
above 1 keV. The spectral parameters agreed with those obtained from
fits to the data which had all flagged events removed. The uncertainties
in the derived parameters are slightly smaller with the relaxed quality
filter but all of the fits have unacceptably high values of
.